FlexStation 3 Benchtop Multi-Mode Microplate Reader User Guide
150
0112-0127 B
2.
Press in the handle at the bottom of the top cover and lift the
cover back off the detection module, revealing the fluidics
module.
Figure 5-7:
Opening the Top Cover
3.
Using the handle, tilt the fluidics module to the left.
Figure 5-8:
Tilting the Fluidics Module
If you can clean up the spill, or remove the tips, without
removing the fluidics module, do so now.
If you need to remove the fluidics module completely, see
Removing the Fluidics Module on page 151
4.
Using the handle, lower the fluidics module carefully back over
the detection module, and make sure it is firmly seated.
5.
Lower the top cover over the instrument.
6.
Reattach the power cord and computer cables.