Troubleshooting Procedures
0112-0127 B
137
Opening the Instrument
1.
Turn the instrument power off and disconnect the power cords
from the instrument and the cables connected to the computer.
2.
Press in the handle at the bottom of the top cover and lift the
cover back off the detection module, revealing the fluidics
module.
Figure 5-3:
Location of Inside Panels and Tip Rack Drawer
3.
Turn the four quarter-turn screws and lift the front panel out.
You might need to inspect the interior of the module to determine if the
pipettor was in the process of a fluid transfer when the power was
interrupted. The pipettor head should be at rest in the uppermost
“home” position. If not, there might be fluid in the tips that needs to be
removed. Instructions for these procedures are in the following topics.
CAUTION!
To avoid damage to the instrument, follow these instructions
and any instruction labels on the instrument exactly.