24
Completion of the cycle
When the pressure in the sterilization chambers returns within the pre-set safety limits, the door lock system is released.
As a consequence, the door status icon
flashes
. At the same time, it also beeps.
1
1 3 4 ° C P O R O U S
C Y C L E C O M P L E T E
8 6 . 2 ° C
- 0 . 0 2 b a r 2 9 : 4 0
The icon for the sterilization process status
is
steady on
.
Open the door and retrieve the sterilized material, using the extractor provided.
The icon
symbol
goes off
.
e
When the door is opened, the report for the sterilization cycle executed is automatically produced (if the printer is installed and selected).
Check and sign this report, file it in a suitable place. Refer to the
Print report examples
.
e
At the end of the cycle, and up to the opening of the door, the heating elements are off. As a consequence, the device is slowly cooling
regardless of what the STAND-BY mode is
.
Now the equipment is
ready
to perform a
new cycle
. Repeat the procedures explained in the Chapter
“
Selecting the program
” .
e
Whenever the sterilizer's' door is not opened at the end of the cycle, the vacuum pump is periodically activated to remove any traces of
condensate from the sterilization chamber. The display shows
:
F O R C E D V E N T I L A T I O N
t o s t o p
3 5 . 2 ° C
- 0 . 0 2 b a r 2 9 : 4 0
Press
to interrupt ventilation and open the door.
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...