Rev. 1
Page 3-
12
LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
One or both the steam generator cartridges
failed
Check the steam generator cartridge.
Replace the failed cartridge – see card Gr1-19.
Band heater in short-circuit.
Check the band heater.
replace the band heater – see card Gr1-18.
Damage during the equipment transport.
Replace the damaged part.
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...