Rev. 1
Card Gr2-
3
EV3 VALVE / COIL / SHAFT
401#00230000
EV3 ASSEMBLY
A#0BP2910000
EV2 – EV3 ASSEMBLY
A#0BP2830000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card
Gr7-1
);
2. Remove the wiring and the sheathed pipes from the valve;
3. Remove the group EV2/EV3 from the support;
4. Disassembly the group EV3 as necessary (see figure) and replace the
involved part;
5. Remount all parts, restore connections proceeding in reverse order as above;
6. Run a sterilization cycle.