Rev. 1
Page 3-
49
A 254
LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
Replace the water pump - see card Gr3-1
ATM2 TIMEOUT
2
nd
pressure pulse
(from -.80 to 0.00 bar)
not performed within
the preset time
2
nd
pressure pulse not
performed within 3’ timeout
The water pump of the steam generator
does not work properly
Replace the power board - see card Gr1-1
Clean the valve
Incorrect operation of the EV6 valve
Replace the EV6 valve - see card Gr3-1
Fault in the pipe connecting the steam
generator
Replace the pipe
Sterilization chamber overloaded
Advice the operator for properly loading of the sterilization
chamber.
Restore the safety thermostat and repeat the sterilization
cycle.
Intervention of the steam generator safety
thermostat
Replace the safety thermostat - see card Gr4-6
Replace the lower section of the steam generator
- see card Gr4-6
Steam generator clogged
Different distilled water must be used
Fault in the board
Replace the board - see card Gr1-1
CPU board uncalibrated
Send the sterilizer to the Mocom Service department
Check the continuity of the steam generator
cartridge
Replace the failed cartridge - see card Gr1-19
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...