Rev. 1
Page 2-
13
1
st
pressure pulse – PP1
When the pre-set vacuum value is reached, steam is injected and the pressure begins to rise (
1st PRESSURE PULSE
), until the established
value is reached.
1
1 3 4 ° C P O R O U S
1 . P R E S S U R E P U L S E
1 0 8 . 0 ° C
0 . 4 7 b a r 0 3 : 5 8
CHAMBER
Water
tank
DESCRIPTION
STATUSES
(P, t)
VP WP
EV1
(no)
EV2
(nc)
EV3
(nc)
EV4
EV5
(no)
EV6
(nc)
Fan
1
st
pressure pulse
from –0,88 to +0,50
bar
from –0,80 to 0,00 bar
(water entry5” at –0,80
bar)
from 0,00 to +0,50 bar
OFF ON ON OFF OFF OFF ON ON ON
1
st
pressure pulse
from +0,50 to +1,00
bar
from +0,50 to +0,90 bar
from +0,90 to +1,00 bar
OFF ON
ON OFF OFF OFF ON ON
ON