Rev. 1
Card Gr1-
14
CHAMBER HEATER THERMOSTAT
430#00140000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card
Gr7-1
);
2. Unscrew the cap (1) and the screws from the rear frame;
3. Access to the left side of the frame and withdraw the cable up to the probe;
4. Unlock the spring (2) on the insulation layer to access the he probe’s bulb;
5. Loosen the nuts (3) and withdraw the bulb (4) with the thermostat;
6. Mount the new thermostat, restore connections and assembly all items
proceeding in reverse order as above;
7. Run a sterilization cycle.