Rev. 1
Page 2-
19
Process
When the thermodynamic parameters are balanced, the actual sterilization phase of the materials begins (
HOLDING TIME
).
Thanks to continuous monitoring of the thermodynamic parameters and sophisticated management of the plumbing circuit, the pressure and
temperature are maintained
constant
within the limits required by the program. The display shows the following.
A countdown begins of the sterilization time.
1
1 3 4 ° C P O R O U S
H O L D I N G T I M E
1 3 5 . 6 ° C 0 4 : 0 0
2 . 1 6 b a r 1 3 : 5 5
If, for some reason, the sterilization cycle has been interrupted before completion, the icon continues flashing. If, for some reason, the
sterilization cycle has been interrupted before completion, the icon continues flashing
.
In this case, the material cannot be considered sterile and MUST ABSOLUTELY NOT BE USED..
Water
tank
DESCRIPTION
STATUSES
(P, t)
VP WP
EV1
(no)
EV2
(nc)
EV3
(nc)
EV4
EV5
(no)
EV6
(nc)
Fan
Holding time
For the time set by the
program
OFF ON ON OFF OFF OFF ON ON ON
The icon for the sterilization process status
flashes
to indicate that the
treatment of the load is in progress.
e
At the end of the sterilization phase, the icon
remains steady on to
indicate the complete sterilization of the material in the sterilization chamber
.
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...