Rev. 1
Page 3-
16
LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
Check manually the operation of the locking
micro-switch pin.
- Adjust the micro-switch.
- Replace the micro-switch
- see card Gr6-5
.
Check with a tester the 1,25 A fuse
- Replace the fuse (same value).
- Replace the electronic board
- see card Gr1-1
.
Check for the correct fastening of the motor
mounting screws.
- Restore the correct mounting.
- Replace the motor
- see card Gr6-6
.
- Replace the motor pin
- see card Gr6-6
.
Check for the free turning of the fork
bushes.
- Replace the blocked bush
- see card Gr6-3
.
Check that the motor runs normally.
The locking mechanism is step-working.
- Replace the pin
- see card Gr6-6
.
- Replace the motor
- see card Gr6-6
.
Check the 2 A fuse on the power board
- Replace the fuse
- see card Gr1-16
.
Enter the setup, go to SERVICE and then
select the option LOCKING DEVICE in
order to check the door locking mechanism.
- Replace the burned fuse
- see card Gr1-16
.
- Replace the power board
- see card Gr1-1
.
Door gasket replacement by the user.
- Remove and mount properly the gasket.
-
Replace the door gasket - see card Gr6-1
- Check the parabola adjustment.
-
Replace the parabola- see card Gr6-2
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...