Rev. 1
Card Gr1-
10
PT2 PROBE
430#00180000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card
Gr7-1
)
2. Remove the PT2 wiring from the CPU board;
3. Remove the cable marker no. 2 to be used on the new probe;
4. Access to the left side of the machine and withdraw the wiring up to the
thermal device;
5. Remove the nut and PT2 from its seat;
6. Measure the ohm value of the new PT2 and write the value on the next row
of the label attached on the right rail;
7. Mount the new PT2 device, restore connections and assembly all items
proceeding in reverse order as above;
8. Run a sterilization cycle.
cable marker
PT2 cable
PT2
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...