Rev. 1
Card Gr1-
12
PT4 PROBE
430#00180000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card
Gr7-1
);
2. Remove the PT4 wiring from the CPU board;
3. Remove the cable marker no. 4 to be used on the new probe;
4. Access to the left side of the machine and rear side of the chamber;
5. Remove the nut and the PT4, withdraw the probe from the chamber;
6. Shape the new probe as in figure and mount it, restore connections and
assembly all items proceeding in reverse order as above;
7. Switch on the unit;
8. Run a sterilization cycle.
cable marker
PT4 cable
PT4