Rev. 1
Card Gr6-
5
DOOR-CLOSED SWITCH
433#00010000
DOOR-UNLOCKED SWITCH
433#00010000
DOOR-LOCKED SWITCH
433#00040000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the left-side and top covers (see card
Gr7-1
);
2. Remove the wiring of the switch involved;
A) Door-closed
switch;
B) Door-unlocked
switch;
C) Door-locked switch
3. Remove the screws of the bracket and switch involved;
4. Replace the switch, assembly all items proceeding in reverse order as above;
5. Perform a sterilization cycle.
Note
: To release the door-unlocked switch proceed as follows:
-
Enter the
Setup
;
Service
, digit the code “++--++--“, go to
Device
Test
,
Manual
and
Locking Device
, push key START to enable/disable.
-
After the switch replacement, check its correct operation. Remove the jumper
X21 from the CPU board,
push the hidden key
on the keyboard and switch
on the sterilizer. As message “LOCKING DEVICE” appears on the display,
release the hidden key and push the
Start
key to enable the unlocking of the
closing mechanism; exit the Setup, switch off the sterilizer and restore the
jumper X21.
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...