Rev. 1
Card Gr6-
6
GEAR-MOTOR
403#00010000
PIN
232#A03L25K0
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card
Gr7-1
);
2. Empty both reservoirs;
3. Turn on the left side the unit and remove the ratio-motor access plate (1)
from the bottom frame;
4. Remove the motor wiring and the motor itself (2);
5. Block the motor in a vise and remove the pin (3) by using proper tools;
6. Mount the new pin, the motor taking care to couple two bushes, connect the
wiring (white = + / black = -) and mount the small plate (1) proceeding in
reverse order as above;
7. By means of a large screwdriver, push on the pin (B) of the door-unlocked
switch (
do not release it until explicitly indicated)
and turn on the unit;
8. Wait for the end of the Self test, then enter the SETUP mode;
9. Go to SERVICE option, enter the code “++--++--“, then select DEVICE TEST,
MANUAL and LOCKING DEVICE option;
10. Push the
Start
key to enable the locking mechanism, then
release the
pin of
the door-unlocked switch
;
11. Push again the
Start
key to release the locking mechanism;
12. Switch off/on the unit;
13. Perform a sterilization cycle.
1
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...