Rev.1
Page 1-
28
ACTIVATING CONFIGURATION OPTIONS – ADVANCED MENU
Setting the sterilization programs
(PROGRAMMES on the ADVANCED menu)
The program setting and their storing in four pre-set positions is achieved in various steps using several menus in sequence.
Each pre-set position can be associated to a
standard
or user configurable cycle (
CUSTOM
).
Let's look at the two cases separately.
To associate a
standard program
and define several of its parameters, proceed as follows:
1. Select
PROGRAMS
using the
↵
key; the following menu appears:
→
1 s t P R E S E T +
2 n d P R E S E T
↑
3 r d P R E S E T
↓
4 t h P R E S E T -
E X I T
Define the position (
1
,
2
,
3
or
4
) to which the sterilization program will be associated using the
+
and
-
keys. Confirm with the
↵
key.
2. From here, you enter the list of available cycles:
→
1 3 4 ° C H O L L O W +
1 2 1 ° C H O L L O W
↑
1 3 4 ° C S O L I D
↓
1 2 1 ° C S O L I D -
1 3 4 ° C E M E R G E N C Y
1 3 4 ° C W R A P P E D
1 2 1 ° C W R A P P E D
1 3 4 ° C P O R O U S
1 2 1 ° C P R O R O U S
1 3 4 ° C P R I O N
X X X ° C C U S T O M
E X I T
Using the
+
and
-
keys, scroll the list until you identify the sterilization program desired.
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...