Setting the Water Draining mode
(DRAIN OPTIONS from the ADVANCED menu)
The water used for the sterilization cycle can be collected into either the
internal
tank (standard configuration) or an
external
tank of greater
capacity (offered as an option – see chapter “Installation”) so as to reduce the frequency of used water draining.
After
DRAIN OPTIONS
is selected, the following menu appears:
→
I N T E R N A L D R A I N +
E X T E R N A L D R A I N
↑
E X I T
↓
-
When
INTERNAL DRAIN
is enabled, the reading of the MAX level sensor in the internal tank is enabled.
The
EXTERNAL DRAIN
command also activates the MAX level sensor located in the external tank.
e
The level sensor in the internal tank remains active in any case, to prevent a possible malfunction of the external tank or a missing or
faulty connection of the optional external drain tank.
e
In the case of an installation with the drain connected to the central system, select INTERNAL DRAIN.
Scroll through the items with the
+
and
-
keys; confirm with the
↵
key.