Rev. 1
Page 3-
6
which is automatically transformed to the message:
( A l a r m M e s s a g e )
R E S E T S Y S T E M
8 0 . 5 ° C X X X X
0 . 0 0 b a r
Perform a
RESET
(described below) and then turn-off the device and check the alarm.
e
Alarms that intervene outside of a program do not produce a printed report
.
RESETTING THE SYSTEM
The system is
RESET
in two
alternative
ways, depending on the alarm that occurred:
1. Press
the
PROGRAM SELECTION
key for about 3 seconds.
A beep confirms the RESET;
Never turn the device off before executing a RESET.
2.
Turn-off the device and then power-on using the main switch
.
Upon power-up, the sterilizer will perform its normal initial test
.
After RESET, and any technical intervention necessary to eliminate the fault, the device will go to STAND-BY mode, ready to execute a new
program.