Select item PRINTOUT MODE to chose the mode the data are printed: The following options appear:
→
A T C Y C L E E N D +
S T E P B Y S T E P
↑
E X I T
↓
-
Select
AT CYCLE END
to print the report al the end of the cycle.
Select
STEP BY STEP
to print the data at each phase of the cycle, as result in the normal printout (see
Printed report examples
).
e
The print of the Vacuum and Helix Test report is carried out only in mode “At Cycle End”.
Activate
NR. COPIES
to set the number of copies of the cycle report to print at the end of the program. The following text appears:
C O P I E S : X X
+ / - t o s e t
↵
t o e n t e r
↑
t o e x i t
Set the number of copies desired (up to a maximum of
5
). Confirm with the
↵
key.
e
In STEP BY STEP mode is not possible more report copies.
On the other hand, the selection
LAST
reprints the report for the last cycle executed (whether it terminated correctly or was
interrupted by an alarm). The following screen appears:
→
N O R M A L P R I N T +
E X T E N D E D P R I N T
↑
E X I T
↓
-
The
NORMAL PRINT
command activates normal printing (that with salient cycle data produced at the end of a correctly executed cycle),
while
EXTENDED PRINT
activates complete printing (including all the data typical of a cycle interrupted by an alarm).
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...