THE PROGRAMS AVAILABLE
GENERAL
The steam sterilizer is appropriate for almost all materials and instruments, so long as they are able to tolerate, without damage, a
minimum
temperature of 121 °C
(otherwise, you will need to use other low-temperature sterilization systems).
The following material can normally be sterilized with steam:
– Stainless steel surgical instruments;
– Carbon steel surgical/generic instruments;
– Rotating and/or vibrating instruments driven by compressed air (turbines) or mechanical transmission (counter-angles, tooth scalers, etc.);
– Glass
articles;
– Mineral-based
articles;
– Articles made of heat-resistant plastic;
– Articles made of heat-resistant rubber;
– Heat-resistant
textiles;
– Medication materials (gauze, pads, etc.);
– Other generic material suitable for autoclave treatment.
e
Depending on the conformation of the material (solid, hollow or porous), any packaging (paper/plastic envelope, sterilization paper,
container, muslin napkin, etc.) and its heat-resistance, it is indispensable that you choose the appropriate program (see
Program
summary table
)
.
The device must not be used for sterilizing fluids, liquids or pharmaceutical products.
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...