Rev.1
Page 1-
32
Select
FRACTION.
to perform a fractionated vacuum (indispensable for sterilizing hollow bodies and porous materials), or
SINGLE
for a
single preliminary vacuum phase (solid instruments). Move using the
+
and
-
keys and confirm with the
↵
key.
4. At this point, you come to another menu where you set the drying mode:
→
S H O R T D R Y I N G +
L O N G D R Y I N G
↑
E X I T
↓
-
Select
LONG
drying suitable for porous and/or wrapped loads, or
SHORT
if you need to sterilize solid, loose materials (and even hollow so
long as not wrapped). Move with the
+
and
-
, confirm with the
↵
key.
5. Depending on the selection (
SHORT
or
LONG
) one of two different menus will open (these menus are the same for the standard cycles), i.e..
In
SHORT
mode the following is displayed:
→
S T A N D A R D D R Y I N G +
F A S T D R Y I N G
↑
E X I T
↓
-
In
LONG
mode the following is displayed :
→
S T A N D A R D D R Y I N G +
I N T E L L . D R Y I N G
↑
E X T R A D R Y I N G
↓
E X I T -
For the choice criteria, refer to the instruction.
Whenever the CUSTOM program is already present in another position, the selection is not accepted. The following warning appears on the display, along
with a beep:
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...