LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
LOCKING
PROBLEM
Failure on the micro-
switches (OFF-OFF)
During the initial self test the
two micro-switches are ON
At the switching on with door open, the
micro-switch is broken.
Replace the micro-switch - see card Gr6-5
At the switching on with door open, the
micro-switch wiring is broken.
Replace micro-switch wiring
Release the micro-switch
At the switching on with door open, the
micro-switch is blocked.
Replace the micro-switch - see card Gr6-5
Replace the micro-switch - see card Gr6-5
At the switching on with door closed, the
wheel-pin micro-switch is broken.
Go to Attachment H or Attachment I
Release the micro-switch
Replace the micro-switch - see card Gr6-5
At the switching on with door closed, the
micro-switch is blocked.
Go to Attachment H or Attachment I
Replace the fuse - - see card Gr1-16
Fuse burned during the previous cycle
Go to Attachment H or Attachment I
Go to Attachment H or Attachment I
Alarm occurred during the previous cycle;
sterilizer turned off before completing the
safety procedure
Advice the user for the correct operation in case of alarm
The alarm continues to occur
replace the CPU board – see card Gr1-1.
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...