Rev.1
Page 1-
44
Device Test
(DEVICE TEST item on the SERVICE menu)
Through this option it is possible to check any device of the sterilizer. The following display is shown:
→
A U T O M A T I C +
M A N U A L
↑
E X I T
↓
-
Use key
+
/
-
to select the test mode, push the
↵
key
to confirm.
In
AUTOMATIC
, the unit requires a confirmation and then start the automatic test of the devices according to the following order:
LOCKING DEVICE
WATER PUMP
EV1 VACUUM
PUMP
EV2
AUXILIARY PUMP (automatic water feeding)
EV3 STEAM
GENERATOR
EV4 RESISTOR
(heating)
EV5 ELECTRIC
FANS
EV6
During the automatic test sequence, the display will show
NOW TESTING ….
and at the end of the procedure the message
AUTOMATIC TEST COMPLETE
.
In
MANUAL
mode the display shows the device list; scroll the list and select the item on which the test must be performed.
Use key
+
/
-
to select the device, push the
↵
key
to confirm.