Safety valve maintenance
Access the safety valve located on the rear of the machine.
Loosen the knurled locking ring with your fingers (or a suitable tool inserted in the two holes in the ring itself), turning counter-clockwise until it
reaches the end and turns loosely.
Retighten the locking ring and repeat the operation a couple of times .
Definitively
tighten the locking ring all the way down.
T
HIS OPERATION IS NECESSARY TO GUARANTEE THE CORRECT FUNCTIONING OF THE VALVE OVER TIME
.
A
T THE END OF MAINTENANCE
,
MAKE SURE THAT THE LOCKING RING IS COMPLETELY SCREWED ON AND TIGHTENED
.
Clean (or replace) the water drain filter
With use, various residues will probably tend to accumulate inside the filter, obstructing the lower drain tube over time.
For cleaning (or replace) the filter, open the door of the sterilizer and remove the nut (
1)
with a hexagonal wrench n. 14.
Then remove the fitting (
2)
and the filter (3).
Remove the filter from the support and carefully clean it under a throw of running water, using if necessary a pointed tool to remove possible
material of greater dimensions.
If the filter cannot be reused, replace it with a new one.
Plug the filter in the support, block it with a drop of sealing (if available), having care to not obstruct the holes
Reassemble all parts following reversely the above procedure, paying attention on screwing down the fitting (2) so as to let the draining holes
(4)
at level of the chamber wall.
Summary of Contents for millennium B
Page 1: ...Rev 1...
Page 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Page 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Page 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Page 272: ...Rev 1 Page 4 6 TABLE 2...
Page 273: ...Rev 1 Page 4 7 TABLE 3...
Page 274: ...Rev 1 Page 4 8 TABLE 4...
Page 275: ...Rev 1 Page 4 9 TABLE 5...
Page 276: ...Rev 1 Page 4 10 TABLE 6...