Rev.1
Page 1-
31
To define the
CUSTOM
program to associate to one of the pre-set position (
1
,
2
,
3
or
4
) proceed as follows:
1. Select
PROGRAMS
, select the program number to which the program is to be associated (see the previous description) and then select
CUSTOM
in the next
screen; the following menu appears:
→
1 3 4 ° C P R O C E S S +
1 2 1 ° C P R O C E S S
↑
E X I T
↓
-
Select
121 °C
to perform a custom program with a sterilization process at
121 °C
or
134 °C
for one at 134 °C. Move using the
+
and
-
keys
and confirm with the
↵
key.
2. You will then go the screen:
T I M E : X X m i n
+ / - t o s e t
↵
t o e n t e r
t o e x i t
Use the + and - keys to set the duration of the sterilization process and confirm with the
↵
key.
e
The duration of the sterilization process is variable from 4 to 30 minutes for the program at 134 °C, and from 20 to 30 minutes for the
program at 121 °C.
3. After selecting the time, you go to the menu where you specify the type of initial vacuum:
→
F R A C T I O N . V A C U U M +
S I N G L E V A C U U M
↑
E X I T
↓
-