Rev. 1
Page 2-
14
2
nd
vacuum pulse- PV2
At the end of the pressure rise, the steam, mixed with residual air, is discharged and the second emptying of the sterilization chamber begins
(
2nd VACUUM PULSE
).
1
1 3 4 ° C P O R O U S
2 . V A C U U M P U L S E
9 3 . 3 ° C
- 0 . 7 9 b a r 0 6 : 0 6
CHAMBER
Water
tank
DESCRIPTION
STATUSES
(P, t)
VP WP
EV1
(no)
EV2
(nc)
EV3
(nc)
EV4
EV5
(no)
EV6
(nc)
Fan
Steam discharge
from +1,00 to +0,22 barr
OFF OFF ON OFF ON OFF ON OFF ON
Transition phase of the
charging circuit (1”)
ON OFF ON OFF OFF OFF ON OFF ON
Vacuum of the discharge
circuit
ON OFF ON OFF OFF ON ON OFF ON
Chamber
discharge
ON OFF ON OFF ON ON ON OFF ON
Chamber vacuum (up to -
0,80 bar)
2
nd
vacuum pulse
Chamber vacuum (up to
PT2 =165°)
ON OFF OFF OFF ON ON ON OFF ON