Rev. 1
Card Gr1-
5
PRESSURE TRANSDUCER
431#00080000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
.
1. Remove the covers ((see card
Gr7-1
);
2. Remove the pressure transducer wiring from the CPU board;
3. Withdraw the pipe from the pressure transducer;
4. Remove the nut (1) fixing the pressure transducer to the bracket;
5. Remove the pressure transducer;
6. Mount the new pressure transducer, restore connections and assembly all
items proceeding in reverse order as above;
7. Calibrate the pressure transducer –
see Attachment N
8. Run a sterilization cycle.