Rev. 1
Card Gr7-
1
LEFT-SIDE COVER
C#1BP1660001
TOP COVER
C#1BP1650001
SINGLE COVER
C#1BH0060001
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1.
Remove the screws of the left-side cover (1);
2.
Remove the screws of the top cover (2)
3.
Attach new adhesive strips (3);
4.
Remove the four screws of the single cover (4);
5.
Remount the new covers;
6.
Perform a sterilization cycle.
1
1
2
4
3