4 - 3
Repair and Disassembly
Disassembly
4.3
Disassembly
4.3.1
Separating the Front and Rear Halves of the Monitor
1.
If the monitor is not configured with an Optional Temperature module, proceed to
step 2. If the monitor is configured with an Optional Temperature module:
a.
Remove the two (2) M3×6 cross-head screws (as shown in Figure 4-1).
FIGURE 4-1
Remove two (2) M3x6 cross-head screws.
b.
Disconnect the Temp cable connector from the PCBA as shown in Figure 4-2.
FIGURE 4-2
Disconnect the Temp cable connector.
M3x6 cross-head screw
M3x6 cross-head screw
Temp cable connector
Summary of Contents for Rosebud Accutorr V