Visual Test
Test and Calibration
6 - 4
6.2
Visual Test
Inspect the equipment for obvious signs of damage. Follow these guidelines when inspecting
the equipment:
• Carefully inspect the case, the display screen, and the buttons for physical damage.
• Inspect all external connections for loose connectors, bent pins, or frayed cables.
• Inspect all connectors on the equipment for loose connectors or bent pins.
• Make sure that the labels on the equipment are clearly legible.
6.3
Power-on Test
Verify that the monitor can power up correctly by following this procedure:
1.
Insert the lithium battery in the battery compartment and connect the monitor to the AC
mains. The AC mains LED and battery LED light.
2.
Press the Power On/Off button to turn on the monitor.
• The operating status LED built into the Power On/Off button illuminates.
• The system beeps to indicate it has passed the alarm sounds self test.
• All the LEDs on the front panel illuminate.
• The technical alarm lamp illuminates yellow, then red, and then turns off to indicate it
has passed the alarm lamp self test.
• The start-up screen clears, and the monitor enters the main screen, indicating start-up is
complete.