4 - 17
Repair and Disassembly
Disassembly
4.3.11
Disassembling SpO
2
Modules
4.3.11.1
Disassembling a DPM SpO
2
module
1.
Remove the M3×6 cross-head screw (as shown in FIGURE 4-28).
FIGURE 4-28
Remove one M3 X 6 cross-head screw
2.
Remove the SpO
2
board. (as shown in FIGURE 4-29)
FIGURE 4-29
DPM SpO
2
board and isolated power supply.
M3x6 cross-head screw