NIBP Calibration
Test and Calibration
6 - 6
FIGURE 6-3 NIBP TOOLS
dialog
9.
Once
CALIBRATION
is highlighted, press
(18) to display the Enter Password
dialog, as shown in FIGURE 6-4.
FIGURE 6-4 ENTER PASSWORD
dialog
10.
In the Enter Password dialog, press
(18). See FIGURE 6-4.
11.
Press
(16) or
(19) to set the first password digit to 3.
12.
Press
(18).
13.
Press
(16) or
(19) to highlight the second password digit.
14.
Press
(18).
15.
Press
(16) or
(19) to set the second password digit to 2.
16.
Press
(18).
17.
Press
(16) or
(19) to highlight the third password digit.
18.
Press
(18).
19.
Press
(16) or
(19) to set the third password digit to 1.
20.
Press
(18).
21.
Press
(16) or
(19) to highlight
OK
.
22.
Press
(18).