Hardware Structure
Theory of Operation
1 - 10
1.4.5
Recorder
The recorder receives data from the main board and sends it to the thermal printhead for
printing. It has a button to start or stop printing and a green LED to indicate the presence or
absence of paper.
1.4.6
Bar Code Scanner
The bar code scanner reads one-dimensional and two-dimensional bar codes to simplify
admitting a patient. The bar code scanner communicates with the monitor. The serial port
supplies it with power. The scanner’s serial port is defined in the following table:
The monitor’s serial port is defined in the following table:
Basic settings of the bar code scanner are listed in the following table:
PIN
DEFINITION
2
Barcode_RX
3
Barcode_TX
5
GND
9
VCC5VDC
PIN
DEFINITION
2
Monitor_TX
3
Monitor_RX
5
GND
9
VCC5VDC
HOST PARAMETERS
BAR CODE SCANNER
FACTORY DEFAULT
Baud Rate
9600
9600
Data Bits
8
8
Stop Bits
1
1
Calibration bit
0
0
Handshaking None
None
USER PARAMETERS
BAR CODE SCANNER
FACTORY DEFAULT
Beeper Tone
Medium
Medium
Beeper Volume
Medium
High
Trigger Mode
Level
Auto Aim
Parameter Scanning
Disable
Enable
DATA FORMAT
BAR CODE SCANNER
FACTORY DEFAULT
Prefix Value
7013 <CR><LF>
7013 <CR><LF>
Suffix 1 Value
Suffix 2 Value
7013 <CR><LF>
7013 <CR><LF>
Scan Data Transmission Format <PREFIX><DATA><SUFFIX 1><SUFFIX 2>
Data only
Summary of Contents for Rosebud Accutorr V
Page 1: ...USER MANUAL Vital Signs Monitor...
Page 125: ...0070 00 0702 Rev 7 0 October 2015...