3 - 5
Troubleshooting
Troubleshooting Guide
3.5.4
Alarm Problems
3.5.5
Key Failure
SYMPTOMS
POSSIBLE CAUSE
CORRECTIVE ACTIONS
The alarm lamp does
not light or is
extinguished, but alarm
sounds
Cables defective or
poorly connected
•
Confirm that the cables from alarm LED board to keyboard
and keyboard to main board are properly connected.
•
Confirm that the connecting cables and connectors are not
damaged.
Key and display board
failure
Replace the key and display board.
Main board failure
Replace the main board.
Alarm does not sound,
but alarm lamp lights
properly
Audio alarm disabled
1.
Press
to display the SYSTEM SETUP dialog as shown
in FIGURE 3-1.
2.
Press
or
to highlight MAINTENANCE.
3.
Once MAINTENANCE is highlighted, press
to
display the MAINTENANCE dialog as shown in FIGURE 3-2.
4.
Press
or
to select USER MAINTENANCE.
5.
Once USER MAINTENANCE is highlighted, press
to
display the USER MAINTENANCE dialog.
6.
Enter the required password.
7.
In the pop-up menu, set the [MIN ALARM VOL] to a value
other than zero.
Cable defective or
poorly connected
•
Confirm that the cable between speaker and main board is
properly connected.
•
Confirm that the connecting cables and connectors are not
damaged.
Audio program error
identified in the system
program
Upgrade system programs.
Speaker failure
Replace the speaker.
Main board failure
Replace the main board,
SYMPTOMS
POSSIBLE CAUSE
CORRECTIVE ACTIONS
Keys do not work
Cables defective or
poorly connected
•
Confirm that the cable between keyboard and main board
is properly connected.
•
Confirm that the connecting cables and connectors are not
damaged.
Keyboard failure
Replace keyboard.