2 - 5
Specifications
NIBP Measurement Cycle
2.3.2
Automatically Adjusted Inflation Value after First Measurement
In the Interval and Stat modes, the inflation value of the next measurement will be
automatically adjusted by the NIBP module based on the systolic value of the previous
reading. After the first unsuccessful measurement, the subsequent inflation value will be the
previous systolic pressure reading plus P2:
2.3.3
Pulse Rate
Accuracy:
±3 BPM or ±3%, whichever is greater.
Resolution:
1 BPM
Adult Mode:
P2 = 50
±
10 mmHg
Pediatric Mode:
P2 = 50
±
10 mmHg
Neonate Mode:
P2 = 40
±
10 mmHg
PATIENT TYPE
PULSE RANGE
Adult
35 – 245 BPM
Pediatric
35 – 245 BPM
Neonate
70 – 245 BPM
Measurements outside of the stated ranges are not
guaranteed to be accurate.