NIBP Accuracy Test
Test and Calibration
6 - 12
FIGURE 6-12
NIBP TOOLS dialog running the leak test
NOTE:
The LEAK TEST button changes to STOP LEAK TEST.
Use
or
to highlight STOP LEAK TEST, then press
to
cancel the leak test.
11.
If the system leaks, the message
PNEUMATIC LEAK
will be displayed in the technical
alarm area/prompt area. In this case, check for a loose connection and perform the test
again.
NOTE:
If the system does not leak, the Accutorr V does not display
a message.