Temperature
Specifications
2 - 10
2.5
Temperature
2.6
Display Area, Indicator, and Controller
On the front panel of the Accutorr V, there are a panel display area, a red/yellow LED group
(on the top of the device), a set of buttons, and patient cable connectors. Refer to Chapter 2
of the Accutorr V Operating Instructions (P/N 0070-10-0699-XX).
2.6.1
Display Area
The display area includes 1 LCD, 7 groups of 7-segment digit display, and 7 groups of
monochrome LED indicators.
2.6.1.1
LCD
MODE
MEASUREMENT
RANGE
ACCURACY
TYPICAL
MEASUREMENT
TIME
monitor
25
º
C –
44
º
C
(77
º
F – 111.2
º
F)
25
º
C (77
º
F) – 32
º
C
(89.6
º
F), not including
32
º
C (89.6
º
F)
±0.2
º
C
(±0.4
º
F)
<60 s
32
º
C (89.6
º
F) – 44
º
C
(111.2
º
F), including
32
º
C (89.6
º
F)
±0.1
º
C
(±0.2
º
F)
predict
35
º
C –
43
º
C
(95
º
F – 109.4
º
F)
Not defined
typically 10 s – 12 s
from the moment
when temperature
“----” displays
dynamically
LCD type
SIZE AND
RESOLUTION
DISPLAYED CONTENTS
NOTE
Monochrome FSTN
Matrix: 320×160
dot pitch: 0.24×0.24
Startup screen, menus/
dialogs, time and date,
prompt messages, patient ID,
patient size, trend data
(including scroll bar), SpO
2
waveform, and time to the
next NIBP measurement
The brightness
and contrast of
the LCD are
adjustable.
Color TFT
Matrix: 480×272
dot pitch:
0.16125×0.16125
Startup screen, menus/
dialogs, time and date,
prompt messages, patient ID,
patient size, trend data
(including scroll bar), SpO
2
waveform, and time to the
next NIBP measurement
Only brightness
of LCD is
adjustable.