6 - 7
Test and Calibration
NIBP Calibration
NOTE:
When the NIBP CALIBRATION dialog is first displayed,
INFLATE is highlighted, as shown in FIGURE 6-5.
FIGURE 6-5 NIBP CALIBRATION
dialog with
INFLATE
highlighted
23.
Press
(18) to start pressurizing the calibration vessel.
NOTE:
Use
(16) or
(19) to highlight CANCEL, then press
(18) to cancel the calibration.
NOTE:
After the calibration vessel is pressurized, the reference
pressure field is highlighted, as shown in FIGURE 6-6.
FIGURE 6-6 NIBP CALIBRATION
dialog with the reference pressure field highlighted
24.
Press the
(16) or
(19) to adjust the reference pressure value in 1 mm
increments until it matches the
MEASURED PRESSURE
value.
25.
Press
(18) to set the reference pressure value. After the reference pressure value is
set,
ACCEPT
is highlighted, as shown in FIGURE 6-6.