Troubleshooting Guide
Troubleshooting
3 - 6
3.5.6
Recorder Failures
3.5.7
Interface Failures
3.5.8
Power Supply Failures
SYMPTOMS
POSSIBLE CAUSE
CORRECTIVE ACTION
No printout
Recorder module
disabled
•
Confirm that the recorder status LED is on.
•
If yes, reset the recorder. If no, check for other possible
causes.
Paper reversely installed
Re-install the paper roll.
Cable defective or
poorly connected
•
Confirm that the cable between recorder and main board is
appropriately connected.
•
Confirm that the connecting cables and connectors are not
damaged.
Recorder power supply
failure
Confirm that the power module outputs 5V DC and 16.8V DC
correctly.
Recorder failure
Replace the recorder.
Poor print quality or
paper not feeding
properly
Paper roll not properly
installed
Stop the recorder and re-install the paper roll.
Dirty thermal print head
•
Confirm that the thermal print head and the paper roller do
not contain foreign matter.
•
Clean the thermal print head with an appropriate detergent.
Recorder failure
Replace the recorder.
SYMPTOMS
POSSIBLE CAUSE
CORRECTIVE ACTION
No nurse call signals are
issued
Incorrect function settings
1.
Press
to display the SYSTEM SETUP dialog as
shown in FIGURE 3-1.
2.
Press
or
to highlight MAINTENANCE.
3.
Once MAINTENANCE is highlighted, press
to
display the MAINTENANCE dialog as shown in
FIGURE 3-2.
4.
Press
or
to select NURSE CALL.
5.
Once NURSE CALL is highlighted, press
to
display the NURSE CALL dialog.
6.
Select desired nurse call output.
Main board failure
Replace the main board.
Bar code scanner unable
to be used
Bar code scanner failure
Refer to Bar Code Scanner Failures.
Serial port failure
Replace the main board.
SYMPTOMS
POSSIBLE CAUSE
CORRECTIVE ACTION
Fluctuating battery voltage
Battery failure
Replace battery.
Cable defective or poorly
connected
•
Confirm that the cable between battery interface
board and power module is correctly connected.
•
Confirm that the cables and connectors are not
damaged.
Power board failure
Replace the power board.