3 - 9
Troubleshooting
Troubleshooting Guide
3.5.12
NIBP Measurement Failures
3.5.13
SpO
2
Measurement Failure
3.5.14
Temperature Measurement Failures
SYMPTOMS
POSSIBLE CAUSE
CORRECTIVE ACTIONS
Unable to start NIBP
measurement
Cables detective
Confirm that the cables between NIBP module and the
main board are in good condition.
Main board failure
Replace the main board.
Cuff inflation/deflation
failure
Pump or valve damaged
Replace the NIBP module.
Air tubing occluded
Replace the air tubing.
SYMPTOMS
POSSIBLE CAUSE
CORRECTIVE ACTIONS
Unable to measure SpO
2
SpO
2
sensor failure
Replace the SpO
2
sensor.
Cables defective
Confirm that the cable between the SpO
2
module and
the main board is in good condition.
SpO
2
module failure
Replace the SpO
2
module.
Main board failure
Replace the main board.
SYMPTOMS
POSSIBLE CAUSE
CORRECTIVE ACTIONS
Unable to measure patient
temperature
Temperature probe damaged
Replace the Temperature probe.
Cables detective
Confirm that the cable between the Temperature
module and the main board is in good condition.
Main board failure
Replace the main board.