Nurse Call Performance Test
Test and Calibration
6 - 18
21.
Press
or
to highlight the
SIGNAL DURATION
pull-down list.
22.
Once the
SIGNAL DURATION
pull-down list is highlighted, press
to select it.
23.
Press
or
to highlight the
CONTINUOUS
.
24.
Once the
CONTINUOUS
is highlighted, press
to select it.
25.
In the
NURSE CALL SETUP
dialog, set
SIGNAL TYPE
to
NORMAL OPEN
. Trigger
an alarm and measure the contact output with the multimeter. The output should be
continuous high level.
Contact Technical Support if the nurse call test fails.