4 - 7
Repair and Disassembly
Disassembly
2.
Remove the four (4) M3×8 cross-head screws (as shown in Figure 4-7).The monitor with
color LCD does not need this step.
FIGURE 4-7
Remove four (4) M3x8 cross-head screws.
3.
Disconnect the flexible cables on the button board.
4.
Remove the LCD.
5.
Remove the two (2) M3×6 cross-head screws (as shown in FIGURE 4-8), and take out
the SpO
2
rack.
FIGURE 4-8
Remove two (2) M3x6 cross-head screws.
M3x8 cross-head screw
M3x8 cross-head screw
M3x8 cross-head screw
M3x8 cross-head screw
M3x6 cross-head screw
M3x6 cross-head screw