6 - 9
Test and Calibration
NIBP Accuracy Test
6.5
NIBP Accuracy Test
Required Tools:
• Two (2) T-Connectors
• Five (5) pieces of tubing
• One (1) Metal Vessel with volume of 500 ±25 ml
• One (1) Reference manometer with accuracy of 1 mmHg
• One (1) Ball Pump
To calibrate NIBP:
1.
Attach the calibration vessel, ball pump, and reference manometer as shown in
FIGURE 6-8.
FIGURE 6-8
NIBP accuracy test configuration
2.
If needed, press
to display the Normal Screen.
3.
Press
to display the
SYSTEM SETUP
dialog.
4.
Press
or
to highlight
MAINTENANCE
.
5.
Once
MAINTENANCE
is highlighted, press
to display the
MAINTENANCE
dialog.
6.
Press
or
to highlight
NIBP TOOLS
.
7.
Once
NIBP TOOLS
is highlighted, press
to display the
NIBP TOOLS
dialog as
8.
Press
or
to highlight
ACCURACY TEST
.
9.
Once
ACCURACY TEST
is highlighted, press
to start the
ACCURACY TEST
as
Accutorr V
Calibration
Vessel
Reference
Manometer
Ball Pump
Hose