6 - 5
Test and Calibration
NIBP Calibration
6.4
NIBP Calibration
Required Tools:
• One (1) T-Connector
• Three (3) pieces of Tubing
• One (1) Metal Vessel with volume of 500 ±25 ml
• One (1) Reference manometer with accuracy of 1 mmHg
To calibrate NIBP:
1.
Attach the calibration vessel and reference manometer as shown in FIGURE 6-1.
FIGURE 6-1
NIBP calibration configuration
2.
If needed, press
(11) to display the Normal Screen.
3.
Press
(12) to display the
SYSTEM SETUP
dialog.
4.
Press
(16) or
(19) to highlight
MAINTENANCE
.
5.
Once
MAINTENANCE
is highlighted, press
(18) to display the
MAINTENANCE
dialog as shown in FIGURE 6-2.
FIGURE 6-2 MAINTENANCE
dialog
6.
Press
(16) or
(19) to highlight
NIBP TOOLS
.
7.
Once
NIBP TOOLS
is highlighted, press
(18) to display the
NIBP TOOLS
dialog.
8.
Press
(16) or
(19) to highlight
CALIBRATION
, as shown in FIGURE 6-3.
Accutorr V
Calibration
Vessel
Reference
Manometer
Hose