Hardware Structure
Theory of Operation
1 - 4
1.4
Hardware Structure
FIGURE 1-3 shows the Accutorr V’s Hardware Structure. The core of the system is the main
board, which supplies power for all parameter modules. The parameter modules directly
communicate with the main board. The the main board processes measurements and the
status of all modules, and displays them through the key and display board.
FIGURE 1-3
The Accutorr V’s Hardware Structure
NOTE:
The SpO
2
isolation power board does not apply to models
with DPM Pulse Oximetry.
ACCUTORR V
ISOLATION
POWER BOARD
TEMP MODULE
SpO
2
MODULE
NIBP MODULE
FAN
RECORDER
SPEAKER
KEY &
DISPLAY
BOARD
BATTERY
AC
NURSE CALL
RS232
ETHERNET
MAIN BOARD
POWER BOARD
COMMUNICATIONS
/POWER SUPPLY
ISOLATION
POWER BOARD
POWER
SUPPLY
LED &
KEY
LCD
COMMUNICATIONS
/POWER SUPPLY