Disassembly
Repair and Disassembly
4 - 10
4.
If the device is not configured with an optional Temperature module, skip to Step 6.
5.
If the device is configured with an optional Temperature module, remove the
Temperature module board.
FIGURE 4-14
Remove optional temperature module board.
6.
Remove the six (6) M3×6 cross-head screws (as shown in Figure 4-15).
7.
Take out the main rack assembly.
FIGURE 4-15
Remove six (6) M3x6 cross-head screws.
M3x6 cross-head screw
M3x6 cross-head screw
M3x6 cross-head screw
M3x6 cross-head screw
M3x6 cross-head screw
M3x6 cross-head screw