Introduction
Test and Calibration
6 - 2
6.1
Introduction
To ensure the monitor always functions properly, qualified personnel should perform regular
inspections, maintenance, and testing. This chapter provides the test procedures for the
monitor and includes recommended test equipment and frequency. Qualified personnel
should perform the testing and maintenance procedures as required and use appropriate test
equipment.
The testing procedures provided in this chapter are intended to verify that the monitor meets
the performance specifications. If the monitor or a module fails to perform as specified in any
test, repair or replacement is required to correct the problem. If the problem persists, contact
service.
CAUTION:
All tests should be performed by qualified personnel only.
NOTE:
Qualified personnel should acquaint themselves with the
test tools and confirm that test tools and cables are
applicable.
NOTE:
When fields or buttons in a dialog have a line through them,
they are not available for selection or modification.
6.1.1
Test Report
After completing the tests, service personnel should record test results in this table and retain
them for hospital records.
6.1.2
Recommended Frequency
TEST EQUIPMENT
Name
Model/PN
Next Calibration Due Date
TEST RECORD
No.
Test Item
Test Site
Test Results
1
2
CONCLUSION
Pass/Fail:
Tested by:
Date:
CHECK/MAINTENANCE ITEM
FREQUENCY
Visual test
When first installing or after reinstalling.
Power on test
1. When first installing or after reinstalling.
2. Following any maintenance or replacement of any main unit
parts.
NIBP tests
Accuracy test
1. If the user suspects that the measurement is incorrect.
2. Following any repairs or replacement of the NIBP module.
3. At least once every two years.
Leakage test
Calibration
SpO
2
test
Temperature test