NIBP Accuracy Test
Test and Calibration
6 - 10
FIGURE 6-9
NIBP TOOLS dialog running the accuracy test
NOTE:
The ACCURACY TEST button changes to STOP ACCURACY
TEST.
Use
or
to highlight STOP ACCURACY TEST,
then press
to cancel the accuracy test.
10.
Use the ball pump to pressurize the vessel to each of the three different ranges shown in
following table.
The pressure value in the
MAP LED
display should match the value shown on the
reference manometer within the accuracy for each test range as shown in the table.
11.
Use
or
to highlight
STOP ACCURACY TEST
, then press
to end the
accuracy test.
If the accuracy test failed, send the Accutorr V to service for repair.
6.5.1
NIBP Leakage Test
Required Tools:
• One (1) piece of tubing
• One (1) Cylinder
• One (1) Metal Vessel with volume of 500 ±25 ml
To perform the NIBP leakage test:
1.
Set Patient size to adult by pressing
until Adult is selected. The patient size
changes with each key press.
RANGE
ACCURACY
Low approximately 50 mmHg
±3 mmHg
Medium approximately 150 mmHg
±3 mmHg
High approximately 250 mmHg
±3 mmHg