Tools
Repair and Disassembly
4 - 2
4.1
Tools
During disassembly and replacement, the following tools may be required:
• Phillips screwdrivers
• Wire cutters
• Sharp nose pliers
• Metric wrenches
• Metric nut drivers
4.2
Preparation for Disassembly
Before disassembling the monitor: stop monitoring patients, turn off the monitor, disconnect
all the accessories and peripheral devices, remove the battery, and disconnect the monitor
from the mains power supply.
CAUTION:
Before disassembling the monitor, eliminate static charges.
When disassembling the parts labeled with static-sensitive
symbols, wear electrostatic discharge protection such as
antistatic wristband or gloves. Follow the correct sequence
to disassemble the monitor. Otherwise, the monitor may be
damaged permanently. Disconnect all the cables before
disassembling any parts. Take care not to damage any
cables or connectors.
CAUTION:
Properly connect the cables or wires when reassembling the
monitor to avoid short circuits. When assembling the
monitor, select proper screws. If the wrong size screw is
tightened by force, the monitor may be damaged and the
screw or the part may not function as expected.