NIBP Calibration
Test and Calibration
6 - 8
FIGURE 6-7 NIBP CALIBRATION
dialog with
ACCEPT
highlighted
26.
Press the
(16) or
(19) to select
ACCEPT
.
27.
Press
(18) to accept the calibration.
NOTE:
Use
(16) or
(19) to highlight CANCEL, then press
(18) to cancel the calibration.
The Accutorr V uses the new calibration factor until another calibration sequence is
performed.