6 - 11
Test and Calibration
NIBP Accuracy Test
FIGURE 6-10
Patient size graphics and indicators
The Patient size indicator illuminates to indicate the selected size as shown in FIGURE 6-
10. The factory default setting for the Patient size is Adult.
2.
Connect the metal vessel with the NIBP connector on the Accutorr V monitor.
FIGURE 6-11
Leakage test configuration
3.
If needed, press
to display the Normal Screen.
4.
Press
to display the
SYSTEM SETUP
dialog.
5.
Press
or
to highlight
MAINTENANCE
.
6.
Once
MAINTENANCE
is highlighted, press
to display the
MAINTENANCE
dialog.
7.
Press
or
to highlight
NIBP TOOLS
.
8.
Once
NIBP TOOLS
is highlighted, press
to display the
NIBP TOOLS
dialog as
9.
Press
or
to highlight
LEAK TEST
.
10.
Once
LEAK TEST
is highlighted, press
to start the leak test. The Accutorr V
automatically deflates in approximately 60s, and the NIBP leakage test is complete.
Adult
Pediatric
Neonate
Accutorr V
Hose
Calibration
Vessel